12 INSTRUCTIONS FOR USE
ENGLISH
Chapter 4 Filling with medication
4.1 Pour the medication into the medicine cup:
Open the medicine cup cover.
Fill the medicine cup with the medicine prescribed by your doctor.
The nebulizer has been tested with medicines such as Ventolin and
Pulmicort.
The maximum capacity of the medicine cup is 8 ml.
Close the cover of the medicine cup properly.
Warning:
If the medication solution runs out, tilt the device slightly to ensure
that all the medication is used up.
Suspensions or highly viscous medicines are not suitable for this device.
If sediment builds up on the mesh, nebulization may stop. If this
happens, turn off the power and use gauze or a lint-free towel to
absorb the sediment from the mesh.