21Portable mesh nebulizer YM-3R9
ENGLISH
Chapter 11 Troubleshooting
Refer to the table below if any abnormalities occur during use of the
nebulizer.
Problem Possible causes Solutions
Very little or no mist
comes out of the
inhaler.
The medicine cup is
not attached.
Attach the medicine
cup according to these
instructions, then restart
the device.
The mesh of the med-
icine cup is clogged
with dirt, preventing
atomisation.
Clean the medicine
cup according to these
instructions. Do not use
immediately after cleaning.
If the device still does not
work properly after clean-
ing, replace the medicine
cup with a new one.
There is medication
or water residue on
the top electrode of
the cup.
Remove the medication
residue and rinse according
to these instructions, then
restart the device.
The contact electrode
on the nebulizer and
cup is dirty.
Clean the electrode accord-
ing to these instructions
and restart the unit.
When the device
is switched on, the
ON/OFF indicator
lights up for one
second and then
goes out.
The medicine cup is
not attached.
Attach the medicine
cup according to these
instructions, then restart
the device.
There is no medication
in the medicine cup.
Pour the medicine solution
into the cup according to
these instructions.
The contact electrode
on the nebulizer and
cup is dirty.
Clean the electrode accord-
ing to these instructions
and restart the unit.