23Portable mesh nebulizer YM-3R9
ENGLISH
Problem Possible causes Solutions
Medicine is leaking
from the medicine
cup.
The cup is damaged or
the seal is worn.
Replace the medicine
cup according to these
instructions.
The nebulizer does
not turn off auto-
matically when the
medication runs out.
Some medicines foam
in the cup.
Press the ON/OFF button
to switch off the device
and remove the foam from
the cup.
There is medication
or water residue on
the top electrode of
the cup.
Press the ON/OFF button
to switch off the device
and remove any remaining
medication or water.
The contact electrode
on the nebulizer and
cup is dirty.
Clean the electrode accord-
ing to these instructions
and restart the device.
Damage to the medi-
cine cup.
Replace the medicine
cup according to these
instructions.
Chapter 12 EMC Declaration
1.
The YM-3R9 portable mesh nebulizer is suitable for home and hospital
use, except in areas with active high-frequency surgical equipment
and RF shielded rooms of an ME system for magnetic resonance
imaging, where the intensity of electromagnetic disturbance is high
2.
CAUTION: Portable RF communication devices (including peripheral
devices such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) from any part of the portable mesh
nebulizer, including cables specified by the manufacturer. Failure to
do so may result in reduced performance of this equipment.
3.
IEC 60601-1-2:2014 Identification of MEDICAL ELECTRICAL DEVICES,
marking and documentation of Class B products.