16 INSTRUCTIONS FOR USE
ENGLISH
1.5 To disinfect, immerse the medicine cup in 75% medical alcohol for
at least 3 minutes.
1.6 Remove the medicine cup, wipe off any excess alcohol with a soft,
clean cloth, and leave it to dry.
2.
Step two – cleaning and disinfecting the masks, mouthpiece and
mask connecting tube
2.1 Soak the accessories in clean water (≤50°C) for 2–3 minutes.
2.2 Dry them with medical gauze.
2.3 Wipe accessories with 75% medical alcohol for at least 3 minutes
to disinfect.
CAUTION: If the accessories are not properly cleaned and disinfected,
they may be contaminated with microorganisms that could pose a
risk of infection.
3.
Cleaning the main device (maintenance of the main frame and
cleaning after nebulization)
Clean the main device after use.
3.1 Clean the surface of the main device with a soft, clean cloth damp-
ened with 75% alcohol.
3.2 The electrodes must be cleaned for the device to function properly.
CAUTION:
•
Do not wipe the main device with any other type of volatile liquid,
such as benzene or thinner.
• Do not rinse or immerse the main device in any liquid.
Chapter 9 Technical specification
Device name Portable mesh nebulizer
Model YM-3R9
Dimensions approx. 45×45×100 mm (L)×(W)×(H)
Weight (g) approx. 100 g (without batteries)