15Portable mesh nebulizer YM-3R9
ENGLISH
CAUTION: Batteries included.
CAUTION: Battery life may vary depending on the type of batteries used.
CAUTION: When the red light comes on (indicating low power), replace
both batteries with new ones of the same type.
7. 2 Operation using a mains adapter
7. 2.1. See “3.3.2. Connecting to the mains adapter”, page 10
7.2.2.
Overcurrent and overvoltage protection: The nebulizer detects
an overcurrent or short circuit.
CAUTION: The company does not supply a mains adapter.
• Please use a mains adapter with a DC output voltage
of 5 V, 1 A and conforming to IEC 60601-1, provided
that there are at least two means of patient protection
(MOPP) between the AC input and the DC output.
• Unplug the mains adapter after use. Do not leave the
power cord plugged in when the nebulizer is not in use.
Chapter 8 Cleaning and disinfection
To avoid cross-contamination, each user should have their own set of
net nebulizer accessories. Accessories must be cleaned and disinfected
before use. Switch off the power supply before cleaning and disinfecting.
Detailed cleaning and disinfection procedures are described below:
1. Step one – cleaning the medicine cup
1.1 Open the medicine cup fitted to the main device.
1.2 Pour a small amount (approx. 8 ml) of hot distilled water into the
medication cup and close the cover. The water temperature should
be 30–60°C.
1.3
Press the ON/OFF button, the device will switch on and nebulize
with distilled water for 1–2 minutes.
1.4
Rinse the medicine cup, nozzle attachment and mesh with distilled
water. Gently wipe off excess water with a soft, clean cloth.