18 INSTRUCTIONS FOR USE
ENGLISH
Test with Pulmicort at normal temperature of 24 ±2°C, relative hu-
midity: 45–75% and pressure: 86 kPa to 106 kPa, – for details: chart 1:
Medicines used for testing:
1. For particle size determination - Budesonide (Pulmicort) solution
0.025 % (M/V).
2. For determination of aerosol initial velocity and aerosol efficiency
– Budesonide (Pulmicort) 0.05% (MV) solution.
individual test
mean value
Cumulative percentage (%)
Mean mass fraction (%)
Aerodynamic mean (µm)
Chart 1