25Portable mesh nebulizer YM-3R9
ENGLISH
Magnetic field at
mains frequency IEC
61000-4-8
30 A/m 50 Hz / 60 Hz 30 A/m 50 Hz / 60 Hz
Conduced RF
– IEC61000-4-6
150 KHZ to 80 MHz: 3
Vrms 6Vrms (in ISM and
amateur radio bands)
80% Am at 1 kHz
150 KHZ to 80 MHz: 3
Vrms 6 Vrms (in ISM and
amateur radio bands)
80% Am at 1 kHz
Radiated RF
IEC61000-4-3
10 V/m
80 MHz-2,7 GHz
80% AM at 1 kHz
10V/m
80 MHz-2,7 GHz
80% AM at 1 kHz
CAUTION: U
T
is the AC voltage before the test voltage is applied.
Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
Test Frequen-
cy (Mhz)
Band
(MHz)
Service Modulation Modula-
tion (W)
Distance
(m)
Immunity test
level (V/m)
Radiated radio frequency IEC61000- 4-3 RF wireless commu-
nication devices)
385 380-390 TETRA 400 Pulse
modulation
18 Hz
1,8 0,3 27
450 430-170 GMRS 460, FRS
460
FM ±5 kHz
deviation
1 kHz sine
2 0,3 28
710 704-787 LTE Band 13,17 Pulse
modulation
217 Hz
0,2 0,3 9
745
780
810 800-960 GSM 800/900,
TETRA 800, iDEN
820, CDMA 850,
pasmo LTE 5
Pulse
modulation
18 Hz
2 0,3 28
870
930
1720 1700-1990 GSM 1800;
CDMA1900;
GSM 1900; DECT;
LTE Band 1,3,4,
25; UMTS
Pulse
modulation
217 Hz
2 0,3 28
1845
1970
2450 2400-2750 Bluetooth,
WLAN, 802.11
b/g/n, RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2 0,3 28
5240 5100-5800 WLAN 802.11a/n Pulse
modulation
217 Hz
0,2 0,3 9
5500
5785