Specification
Spec Point Current 10A into 100 mΩ at 240 V. 8,7 A into 100 mΩ at 230 V.
Short Circuit Current Typically 12 A
25A Earth Bond Test
Meter reading range: 0 to 1990 mΩ
Range Resolution Accuracy
0 to 999 mΩ 1 mΩ 2,5% of reading ±5 mΩ
900-1990 mΩ 10 mΩ 2,5% of reading ±10 mΩ
Pass Band Limits: 100 and 500 mΩ
Open Circuit Voltage: Typically 6 V rms a.c. 50 Hz
Spec Point Current: Typically 25A into 100 mΩ (10A into 500 mΩ)
Short Circuit Current Typically 36.5 A
Insulation Test
Meter reading range: 0-99 MΩ
Range Resolution Accuracy
0 to 10 MΩ 0,1 MΩ 2,5% of reading ± 0,1 MΩ
10MΩ to 20 MΩ 0,2 MΩ 5% of reading ± 0,2 MΩ
20MΩ to 50 MΩ 0,5 MΩ 5% of reading ± 0,5 MΩ
50MΩ to 99 MΩ 1 MΩ Indication only
Pass Band limit: 2 MΩ (Class I) – 7 MΩ (Class II)
Open Circuit Voltage: ≤ 600 V d.c.
Spec Point Voltage ≥ 500 V d.c. into 0.5 MΩ
Short Circuit Current: < 2 mA
General
Temperature Range
Operating: 5 ˚C to +40˚ C
Storage: -25˚C to +65˚ C
16