Page 29
Capacitor discharge: 5 kV to 50 V: < 120 ms/ µF
10 kV to 50 V: < 250 ms/ µF
15 kV to 50 V: < 3500 ms/ µF
Capacitance range: With test voltage set above 500V
S1-568 10 nF to 25 µF
S1-1068 10 nF to 25 µF
S1-1568 10 nF to 50 µF
Capacitance accuracy(23 °C):
10 nF to 10 µF : ±10% ±5 nF
Current measurement
range: 0.01 nA to 8 mA
Current measurement
accuracy: ±5% ±0.2 nA at all voltages (23 °C)
Noise rejection:
S1-568: 1 mA per 150 V to a maximum of 8 mA
S1-1068: 1 mA per 320 V to a maximum of 8 mA
S1-1568: 1 mA per 350 V to a maximum of 8 mA
Software filtering: 4 filter settings: 10 s, 30 s, 100 s, 200 s
Voltmeter range: 30 V to 660 V ac or dc, 50/60 Hz
Voltmeter accuracy: ±3%, ±3 V
Frequency range: 45 Hz – 65 Hz
Timer range: 99 m 59 s, 15 s minimum setting
Memory capacity: 11 hrs logging @ 5 sec intervals
Test regimes: IR, IR(t), DAR, PI, SV, DD, ramp test
Interfaces: USB type B (device), Bluetooth
®
Real time output: reading (V, I, R) at a rate of 1 Hz
Remote control: Remote control via USB cable only (requires
RC indicator dongle to be in position)
Environmental Conditions
Altitude: 3000 m (5 kV, 10 kV)
4000 m (15 kV)
Operating temp.: -20 °C to 50 °C
Storage temp.: -25 °C to 65 °C
Humidity: 90% RH non-condensing at 40 °C
Ingress protection: IP65 (lid closed), IP40 (lid open)
General Specifications
Safety: Meets requirements of IEC 61010-1, CATIV 600 V to
3000 m (5 kV, 10 kV)
Meets requirements of IEC 61010-1, CATIV 1000 V
to 4000 m (15 kV)
EMC: Meets the requirements of IEC61326-1
Dimensions:
5 kV, 10 kV L 315 mm x W 285 mm x H 181 mm
15 kV L 360 mm x W 305 mm x H 194 mm
Weight: 4.5 kg (5 kV, 10 kV)
6.5 kg (15 kV)