4 Special function adjustment mode
4.1 User adjustment
4.1.1 Procedure
After the instrument has been switched off, press the [SELECT] key and
hold depressed.
Switch the instrument on using the [ON/OFF] key, after approx. 1 second
release the [SELECT] key.
Press the [MENU] key to select the menu item “Cal”.
Confirm the selection with the [SELECT] key.
The display shows in alternating mode:
Scroll through methods using the [SELECT] key.
Conduct zero setting as described (see section 2.1.1, page 25).
Press the [ZERO/TEST] key.
Method symbol blinks for approx. 3 seconds.
The display shows in alternating mode:
METHOD
METHOD
METHOD
METHOD
CAL
CAL
0.0.0
The Spectroquant
®
Colorimeter Picco features an option that enables user-specific adjust-
ment. This option is required only in exceptional cases.
The Spectroquant
®
rapid tests are subject to stringent checks that ensure a consistently
high quality and a high batch reproducibility. Furthermore, the Picco colorimeters are pro-
grammed by means of calibration at several points.
In the event the user chooses to adjust the instrument, this takes place at only two points
of the measuring range (zero point and measurement result). In this case it is advisable to
check the adjustment using various concentrations of a standard solution.
User adjustment is active:
Factory adjustment is active:
Select
On
Off
On
Off
Select
Zero
Test
Select
Menu
CAL
cAL
Select
Cal