4 Operation – 4.4 Measurements
4.4 Measurements
The spectrophotometer can be used to perform the measurements listed below.
Type of Measurement Description
Concentration Preprogrammed methods that can be executed using Spectroquant® test kits or self-prepared
reagents
User-programmed methods
Absorption / transmission Single-wavelength measurements for establishing the absorbance or transmission of solutions
Multiple-wavelength measurements for establishing the absorbance or transmission of solutions
Spectrum Programmedmethodsforestablishingtheabsorbanceortransmissionofsolutionsoverade󹭯ned
wavelength range
Kinetics Programmedmethodsforestablishingtheabsorbanceortransmissionofsolutionsoverade󹭯ned
period
Quality checks Instrument-supported analytical quality assurance:
Instrument check (AQA 1)
Method-speci󹭯csystemcheck–preprogrammedforallSpectroquant®standards(AQA2)
Pipette volume control (PipetteCheck)
Check interferences from foreign substances (MatrixCheck)
4.3.2 When to repeat the zero adjustment?
We recommend that you repeat the zero adjustment in the
following cases:
If the spectrophotometer was subject to mechanical stress
such as strong shock or transport
If the ambient temperature has changed by more than 5 °C
since the last zero adjustment
At least once a week. The interval to repeat a zero adjustment
is set in the instrument to 7 days. You can change this under
System (Instrument Settings)"
If a new cell type (different manufacturer, different glass type)
is used
Basically, each time you want to measure with the highest
possible accuracy
NOTE
If an interval to repeat a zero adjustment is set you will be
prompted to repeat it after the interval has passed. You can also
repeat a zero adjustment by selecting a method, then touching
the “Settings" icon. Choose "Zero adjustment" and insert a zero
cell to start the measurement.

Other manuals for Merck Spectroquant Prove 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Merck Spectroquant Prove 100 and is the answer not in the manual?

Merck Spectroquant Prove 100 Specifications

General IconGeneral
BrandMerck
ModelSpectroquant Prove 100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals