4.1 Maintenance procedures
We recommend the following advance check-up and maintenance procedures to ensure continued
maintenance and trouble-free operation.
4.1.1 Daily maintenance
• During normal cleaning of the table, make a general visual inspection to ensure that it is in proper
working order.
• The PRACTICO Operating Table must be cleaned after each surgical procedure.
• We recommend that the batteries be charged every day. This will ensure a sufficient battery charge
and long service life
4.1.2 Monthly maintenance
• Lu bricate a ll tabl e jo ints with lig h t mach in e oil .
• Ens u re tha t all cas tors roll eas il y and lock secu re ly.
• Check all table functions by fully extending and retracting them.
• Ensu re that motor-equipped back sections are prope rly m o unte d to the seat. If n ecessary, tighten the
locking screws. See section 4.9 page 26.
• Ensure that the accessory rails are properly mounted.
4.1.3 Annual maintenance
• Perform all the same inspections as in monthly maintenance.
• Check and adjust colum n cle arance, if required.
• Check cables and their connections.
All electrical repairs must be performed by a licensed electrician.
Befo re pe rform in g ma in te na n ce proce d ure s :
• Disconnect power supply
• Turn switch to STAND BY position
• Support the table part being repaired before removing the motor or gas spring.