EasyManua.ls Logo

Merivaara PRACTICO - 4.2 Troubleshooting and Checklist

Merivaara PRACTICO
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
4.2 Troubleshooting and checklist
W hen troubleshooting , check the following items first:
Does the malfunction affect all table functions?
Does the m alfun ction only affect one control function?
If the p robl em on ly affe cts one control fu nction , d oes it do so in b oth di rection s ?
Is the m alfu n ction that the ta ble doe s n ot m ove or that it will no t main tai n its po si tion?
Th e followin g comp one nts a re common to all systems. If there is a malfunction in these, all control functions
will b e affe cted .
Control unit
Hand control unit
Battery
Power source/battery
If there is a malfunction in one of the following components, it usually affects only one function.
Motor
Gas spring
Malfunction Cause
Table adjustments do not function. Sw itch is in STAND BY position
Hand control is unplugged
Battery is dead
Fuse is blown
Faulty ha nd control
Fau lty control unit
Table functions are sluggish and weak. Low battery charge
•Faulty motor
Operating table function duration is short, even
though the battery has bee n charged in accord ance
with instructions.
Fau lty charger
Faulty battery
B atter y too old
Cas tors ma k e ab n orm al nois e s wh en rolli ng Pedal in wron g po si tion
Castor bearings worn
Incorrect castor adj u stm e nt