50
STANDARD COMPLIANCE
Immunity Test IEC 60601
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
according to IEC
61000-4-2
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
-
Test Condition
± 8 kV Contact
± 2, 4, 8, 15 kV Air
± 8 kV contact
discharge
± 2, 4, 8, 15 kV air
discharge
Exempt from this
requirement - cable
length is less than
3 m
MESI mTABLET ABI is intended for use in the electromagnetic
environment specified below. The customer or the user of the
above listed models should assure that they are used in such an
environment.
Patient coupling PORT
Conducted RF
induced by RF
fields
IEC 61000-4-6 (a)
3 Vrms 150 kHz - 80 MHz 6
Vrms in ISM bands between
150 kHz - 80 MHz c)
80% AM 1kHz
Comment: *Not applicable, unit, without Signal input/output parts PORT
a) SIP/SOPS whose maximum cable length is less than 3 m in length are excluded.
b) This test applies only to output lines intended to connect directly to outdoor cables.
c) The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz
to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to
21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.