STANDARD COMPLIANCE
MESI mTABLET ABI is intended for use in the electromagnetic
environment specified below. The customer or the user of the
above listed models should assure that they are used in such an
environment.
Immunity Test IEC 60601
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Test Condition
±,8 kV Contact
± 2, 4, 8, 15 kV Air
±8 kV Contact
± 15 kV Air*
3 V/m
80 MHz – 2.7 GHz
Enclosure Port
Radiated RF EM
fields and
Proximity
fields from RF
wireless
communications
equipment
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80% AM 1 kHz
Mains power quality
should be that of a
professional healthcare
facility environment
and Home healthcare
environment.
27 V/m
385 MHz (18Hz Pulse
Modulation)
450 MHz (FM+/-5 KHz
deviation 1 kHz sine or 18 Hz
Pulse Modulation)
28 V/m
710 MHz (217 Hz PM)
9 V/m
745 MHz (217 Hz PM)
9 V/m
780 MHz (217 Hz PM)
9 V/m
810 MHz (18 Hz PM)
28 V/m
870 MHz (18 Hz PM)
28 V/m
930 MHz (18 Hz PM)
28 V/m
1720 MHz (217 Hz PM)
28 V/m
1845 MHz (217 Hz PM)
28 V/m
1970 MHz (217 Hz PM)
28 V/m
2450 MHz (217 Hz PM)
28 V/m
5240 MHz (217 Hz PM)
9 V/m
5500 MHz (217 Hz PM)
9 V/m
5785 MHz (217 Hz PM)
9 V/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
RATED power
frequency
magnetic fields
IEC 61000-4-8
50 Hz or 60 Hz 30 A/m