EasyManua.ls Logo

METER TEROS 21 - TEROS 21 Measurement Accuracy

METER TEROS 21
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
SYSTEM
3.3.3 MEASUREMENT ACCURACY
TEROS 21 is calibrated at a saturated state (0 kPa), at an air-dry state (−100,000 kPa), and at
three calibration points between 0 and −100 kPa, resulting in accuracy of ±(10% of reading +
2 kPa) over the range of −9 to −100 kPa.
At water potentials drier than −100 kPa, TEROS 21 relies on the linear relationship between
the logarithm of water content and the logarithm of water potential. Laboratory evaluations
have shown good accuracy and low sensor-to-sensor variability to at least −1,500 kPa (plant
permanent wilting point). Field evaluations have shown low sensor-to-sensor variability to
−2,000 kPa (Figure5 and Figure6).
NOTE: METER strongly discourages dry-end calibrations in the pressure plate apparatus. Early attempts to improve
sensor dry-end performance in the pressure plate apparatus actually decreased accuracy, likely because of pressure
plate dry-end equilibrium issues pointed out in the literature (e.g., Campbell [1998], Gee et al. [2002], Bittelli and Flury
[2009], and Frydman and Baker [2009]).
−2,000.0
−1,500.0
−1,000.0
−500.0
0.0
SOIL WATER POTENTIAL (kPa)
TIME (DAYS)
0
20 40
60 80
Sensor 1
Sensor 2
Sensor 3
Permanant wilting point
 Time series TEROS 21 water potential data collected at 80 cm depth
under a beech forest in Switzerland (Walthert, 2013). Note the good sensor


Other manuals for METER TEROS 21

Related product manuals