MI 2094 CE MultiTester General presentation
5
1. General presentation
1.2. Warnings and notes
In order to reach high level of operator’s safety while carrying out various tests and
measurements using CE Multitester, as well as to keep the test equipment undamaged,
it is necessary to consider the following general warnings:
If the test equipment is used in a manner not specified in this user manual the
protection provided by the equipment might be impaired!
Read this instruction manual carefully, otherwise use of the instrument may be
dangerous for the operator, for the instrument or for equipment under test!
Use only grounded mains outlets to supply the instrument!
Do not use any damaged mains outlet or damaged mains connection cable!
A competent authorized person must only carry out Service or calibration
procedure!
Only a skilled person, who is familiar with hazardous voltage operations, can
handle CE MultiTester!
Disconnect all unused test leads before starting measurement; otherwise the
instrument can be damaged!
In case of failure of the instrument test results may be impaired. It is a good
practice to regularly confirm/check the accuracy of the tester (by simple field
calibration, comparison to known reference values etc).
Meaning of , signs on front panel:
Insulation section
Dangerous voltage may be present; measurements
shall be carried out only on de-energized object
Continuity section
Measurements shall be carried out only on de-
energized object. Read instructions on how to
replace blown-up fuse.
Withstanding section
Dangerous voltage may be present. Switch off the
instrument immediately if the red warning lamp (fig.
1., pos. 12) does not light after switching on HV
generator, and service the instrument. Always
handle as the test leads are energized.