EasyManua.ls Logo

Metrix MX 56C - Technical Specifications; DC Voltages; AC Voltages (AC and AC+DC)

Metrix MX 56C
81 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter II
Portable digital multimeter
31
5. TECHNICAL SPECIFICATIONS
Only those values assigned tolerances or limits are guaranteed values. Values without
tolerances are given for information only (French standard NF C 42-670).
{Accuracy : ‘’n% R + nD’’ means ‘’n% of the reading + n digits’’ as per IEC 485}
5.1. DC voltages
Selector
switch position
Ranges Accuracy
Input
impedance
Protection Resolution
mV 500 mV 0.025%R** + 2D 10 M/1G* ± 1100 V
PK
*** 10 µV
5 V 0.025%R** + 2D 11 M ± 1100 V
PK
100 µV
V
DC
50 V 0.025%R** + 2D 10 M ± 1100 V
PK
1 mV
500 V 0.025%R** + 2D 10 M ± 1100 V
PK
10 mV
1000 V 0.025%R** + 2D 10 M ± 1100 V
PK
100 mV
* Refer to § 3.4. ** at 23°C ± 2°C *** 1mn max
Number of points : 50 000 (or 5 000 - refer to § 3.4.)
Range selection : automatic or manual for the 5 V, 50 V, 500 V, 1000 V ranges
Common mode rejection : at 50 and 60 Hz, better than 120 dB
Serial mode rejection : at 50 and 60 Hz, better than 60 dB
Additional error in Pk +/- mode for a pulse of 1 ms : 1% R ± 50 D
For measurements performed on alternative signals, the selected range must tally with the
max. value of the signal peak.
5.2. AC voltages (AC and AC+DC)
Selector
switch
position
Ranges
Accuracy
Input
impedance
Protection Reso-
lution
DC* 40 Hz
to 1 kHz
1 kHz to
4 kHz
4 kHz to
10 kHz
10 kHz to
30 kHz
30 kHz to
50 kHz
from 5 % to 100 % of the range from 10 % to 100 % of the range
mV +
SEL/ON
500 mV* 7 % R
+ 30 D /////////////
10
M/1G**
// 100 pF
±1100V
P
K
***
10 µV
5 V 11 M
// 100 pF
± 1100
V
PK
100
µV
V
AC
or 50 V 0.3% R + 30 D 1 % R
+ 30 D
2 % R + 30 D 3 % R
+ 30 D
10 M
// 100 pF
± 1100
V
PK
1 mV
V
DC
+ 500 V 10 M
// 100 pF
± 1100
V
PK
10 mV
SEL/ON 750 V /////////// ///////////// ///////////// ///////////// 10 M
// 100 pF
± 1100
V
PK
100
mV
* AC+DC only *** 1 mn max ** Refer to § 3.4.
Number of points : 50 000 (or 5 000 - refer to § 3.4.)
Range selection : automatic or manual for the 5 V, 50 V, 500 V, 750 V ranges
Common mode rejection : at 50 and 60 Hz, better than 80 dB
Additional error according to crest factor :
0.2 % for a crest factor of 2 to 3
0.5 % for a crest factor of 3 to 6
(Specification given full scale for a squarewave signal pulse 200 µs wide)