2 - REFERENCE CONDITIONS
Temperature: 23°C ± 2K
Humidity
: 45% RH ± 5%
Position: horizontal ±
Ensure that the pointer is at zero before starting any measurements.
Zero adjustment: open the device. Mechanical zero adjustment is carried out by
turning the transparent cover on the back of the null meter.
Ensure that the switch is correctly positioned.
When an estimated measurement is unknown, place the switch at the highest calibre,
then gradually reduce it until the appropriate calibre is achieved: the reading should
preferably be in the upper 2/3 of the range.
3 - SPECIFICATIONS
3-1 DC VOLTAGE
Connect the leads to the multimeter (be careful of switch position, see
below) and connect to the circuit to be controlled.
When an estimated measurement is unknown, place the switch at the
highest calibre, then gradually reduce it until the appropriate calibre is
achieved.
To obtain voltage in V, multiply the appropriate range value by the reading
coefficient indicated in the table.
V DC
150mV
(1)
0.5V 1.5V 5V 15V 50V 150V 500V 1500V
Range
(indication)
15 (a) 50 (b) 15 (a) 50 (b) 15 (a) 50 (b) 15 (a) 50 (b) 15 (a)
Reading
coefficient
x 10 x 0.01 x 0.1 x 0.1 x 1 x 1 x 10 x 10 x 100
Internal
resistance
(2)
3k 10k 30k 100k 300k 1M 3M 10M 30M
Accuracy
(3)
2%
Admissible
overload
440V
(fuse)
320V (4)
440V (5)
500V 1000V 1500V
(1) Inputs between "COM" and "A" terminals, the other calibres between "COM" and "V " terminals.
(2) specific R: 20k/V
(3) in % of end of range
(4) for 1 minute
(5) for 20 seconds
12