3-4 RESISTANCE MEASUREMENT
Zero adjustment on the ohmmeter is carried out using the calibration button (front) by
short-circuiting the inputs.
x 1 (1) x 10 x 100
Scale (indication) 20k…0 (c)
Reading coefficient x 1 x 10 x 100
Measuring range
10 to 20k 100 to 200k 1k to 2M
Internal resistance
200 2k 20k
End of scale current 8mA 800µA 80µA
Open circuit voltage 1.5V
Accuracy ±10% (2) (4)
Admissible overload 400V (3)
(1) Buzzer for R 100, no buzzer for R 250
(2) of the mid-scale value
(3) for 5 seconds (protection by resistors and CTP)
(4) for battery voltage of 1.45V ± 0.1V
3-4-1 AUDIBLE CONTINUITY TEST - SEMI-CONDUCTOR TEST
NB : In x 1: "buzzer" threshold 250, control of diode on or blocked ( , anode
in "COM" for on). Do not carry out live measurements.
On the ohmmeter, calibre x 1, if voltage of V
AC
> 5V (rms) is applied, the "buzzer"
issues a modulated sound until the error has been corrected.
4 - GENERAL CHARACTERISTICS
4-1 Dimensions and weight
97 x 155 x 43 mm
420 g
4-2 Power supply
1.5V battery (R6 saline or LR6 alkaline)
Battery life:
about 250 hours for the ohmmeter with an alkaline battery
4000 5-second measurements for R < 50 (with beep)
4-3 Maximum climatic conditions
Temperature use -10°C to +50°C ; storage -30°C to +70°C
Relative humidity use 80% HR
Altitude use < 2000m
15