EasyManuals Logo

Metrohm 930 Compact IC Flex ChS/PP/Deg Short Instructions

Metrohm 930 Compact IC Flex ChS/PP/Deg
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
■■■■■■■■
35
Connecting the guard column
Rinsing the guard column
1
Rinsing the guard column
Place a beaker under the guard column's outlet.
Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: in accordance with column leaflet
On
Rinse the guard column with eluent for approx. 5 minutes.
Stop the high-pressure pump in the manual control in MagIC Net
again: Off.
4.22 Connecting the separation column
CAUTION
New separation columns are filled with a solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow the information
provided by the manufacturer).
NOTE
Connect the separation column only after the initial start-up of the
instrument. Until that point, insert a coupling (6.2744.040) instead of
the guard column and separation column.

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex ChS/PP/Deg

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex ChS/PP/Deg and is the answer not in the manual?

Metrohm 930 Compact IC Flex ChS/PP/Deg Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex ChS/PP/Deg
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals