EasyManua.ls Logo

Mettler Toledo IND360 - Page 20

Mettler Toledo IND360
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-16
METTLER TOLEDO IND360fill/dose Application EtherNet/IP Engineering Notes
| 00 | 04/20212
Contents
1 for an example of an AOI instance name that does not match the AOI
name but is still valid.
Q: How do I know the source of the error in the SAI_IND360_FillDose AOI?
A: Typical errors in this AOI include:
o Logic Error: Remember that the SV values must meet the criterion:
HighLimitAlarmSV > HighLimitSV > LowLimitSV > LowLimitAlarmSV.
If this criterion is not met, any execution of a ReadConfig or
WriteConfig will fail until the issue is fixed.
o Fill/Dose PAC is not enabled in the IND360: Using either the local
display or the web interface, access Application > PAC Management in
the IND360 menu system to enable to application.
Q: An AOI is very close to what I want to do in my PLC logic, but I need to
make a few changes. How can I do that?
A: If it necessary to view or modify the logic of an AOI, simply use the
Controller Organizer view in Studio 5000. Navigate to Add-On
Instructions, expand the AOI you are interested in viewing, and double-
click Logic. The Organizer view will show the ladder logic used in the
AOI, and the logic can be changed as necessary for your particular
application.
Figure 5-2: Example of AOI ladder logic

Other manuals for Mettler Toledo IND360

Related product manuals