SF52 User’s Manual
4 97224 Issue 5, June 2017
INSTALLATION
2.3 Transmitter Mounting
Take care to prevent any contamination of the sensor before installation (handle the
transmitter by the main body only, avoiding contact with the lter disc).
The SF52 can be mounted either into a ow-through sensor sampling block (optional)
or directly into a pipe or duct. It can be operated at pressures of up to 2 MPa (20 barg
/ 290 psig) when tted with the bonded seal provided.
The recommended gas ow rate, when mounted in the optional sampling block, is 1 to
5 Nl/min (2.1 to 10.6 scfh). However, for direct insertion applications, gas ow can be
from static to 10 m/sec (32.8 fps).
NOTE: Pass the seal over the mounting thread and assemble into the sampling
location, by hand, using the wrench ats only.
When installed, fully tighten using a wrench until the seal is fully compressed and to the
following torque settings:
G ½” BSP 56 Nm (41.3 ft-lbs)
½” NPT 56 Nm (41.3 ft-lbs)