Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The MicroLab is intended for use in the electromagnetic environment specified below. The
customer or the user of the MicroLab should assure that it is used in such an environment
Electromagnetic Environment -
Guidance
The MicroLab uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
The MicroLab is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The MicroLab is intended for use in the electromagnetic environment specified below. The
customer or the user of the MicroLab should assure that it is used in such an environment.
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
Contact: +/- 8 kV
Air: +/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV
Contact: +/- 8 kV
Air: +/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast
transient / burst
IEC61000-4-4
+/- 2 kV 100 kHz
repetition
frequency for
power supply
lines
+/- 2 kV 100 kHz
repetition
frequency for
power supply
lines
Mains power quality should
be that of a typical
commercial or hospital
environment
+/- 0,5 kV, +/- 1
kV, +/- 2 kV
+/- 0,5 kV, +/- 1
kV, +/- 2 kV
Mains power quality should
be that of a typical
commercial or hospital
environment
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC61000-4-11
0% U
T
; 0,5 cycle
At 0
o
, 45
o
, 90
o
,
135
o
, 180
o
, 225
o
,
270
o
and 315
o
0% U
T
; 1 cycle
and 70% U
T
;
25/30 cycles
Single phase: at
0
o
and 0% U
T
;
250/300 Cycle
0% U
T
; 0,5 cycle
At 0
o
, 45
o
, 90
o
,
135
o
, 180
o
, 225
o
,
270
o
and 315
o
0% U
T
; 1 cycle
and 70% U
T
;
25/30 cycles
Single phase: at
0
o
and 0% U
T
;
250/300 Cycle
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the MicroLab requires
continued operation during
power mains interruptions,
it is recommended that the
MicroLab be powered from
an uninterruptible power
supply or a battery