Page 29
Installation and Assembly
wireSENSOR, WPS
4.3 Wire Guide and Fastening
If the measuring wire has to be extracted from the sensor
to guide the wire respectively to fix it to the target
- the sensor may not be held by another person
- the measuring wire may not be further extracted but
only to the specified measuring range
- the surroundings of the sensor have to be protected
against snapping of the measuring wire
Wrong
Correct
Fix the measuring wire to the target using a wire clip.
Guide the measuring wire vertically out of the sensor
housing.
Misalignment only permissible up to 3 degrees.
Dragging of the measuring wire on the inlet hole or other
objects leads to damage and/or breakage of the measu-
ring wire.
i
If the measuring wire cannot be fed vertically out
of the housing, it is essential to use a guide pulley
(accessory TR1-WDS or TR3-WDS, see Chapter Ac-
cessories).
Keep measuring wire in an area where it cannot be
snagged or otherwise be violated.
max. 3 °
Wire outlet 0 °
±3 ° tolerancy
Fig. 13 Wire fastening and mis-
alignment
A measuring wire under
tension where opera-
tors are standing can
lead to injuries.
Do not twist the