5LIT-PAL-CS Rev B
Cleaning and Processing
Refer to the PAL-650 Instructions for Use (IFU) for cleaning parameters, as any deviation from these
parameters may result in inadequate cleaning of PAL Handpiece and Cable.
Manual Cleaning
1. Thoroughly clean the Handpiece using a bristled brush with warm water (≥49° C / 120° F) then wipe with
a wetted cloth.
2. Apply the Washer Disinfector Cap to the Handpiece or the cable caps to the Cable.
3. Brush instruments again with pH neutral enzymatic detergent. Scrub the Handpiece or Cable with the
brush paying close attention to instrument crevices and connections.
4. Wipe the Handpiece or Cable with a lint-free cloth wetted with warm water (≥49° C / 120° F) until no
visible soil or contaminants remain.
Automated Cleaning
1. Clean the Handpiece or Cable thoroughly using a bristled brush with tap water (30°-40° C / 86°-104° F)
then with pH neutral enzymatic detergent. Pay close attention to instrument crevices and connections.
2. Apply the Washer Disinfector Cap to the Handpiece or the cable caps to the Cable.
3. Load the device into the washer-disinfector. Recommended Washer/Disinfector Cycle is in the PAL
System IFU “IM-PAL-650”.
For more information, please reference the PAL System IFU “IM-PAL-650”.