Steam Sterilization
1. All sterilization cycles require a MINIMUM DRY CYCLE exposure time of 20 MINUTES.
2. This is a minimum requirement and should be increased accordingly until
ALL evidence of moisture has been removed from the device and packaging.
3. Never shorten the dry cycle, this may result in service interruptions related to
residual moisture and corrosion that may not be visible during examination.
4. Always allow the PAL Handpiece and Cable to cool slowly at room temperature.
Attempting use prior to cooling may result in lost functionality and/or damage
the system.
ALWAYS RUN A COMPLETE DRY CYCLE
Gravity-Displacement Steam Sterilization:
Full cycle with 35-minute exposure time at a minimum temperature of 132°C (270°F ), 20 minute minimum
heated dry time.
Cycle Type Gravity-Displacement
Set Point Temperature 132° C / 270° F
Exposure Time 35 minutes*
Dry Time 20 minutes
*Note: This cycle is not a typical cycle as recognized by AAMI ST79.
For more information, please reference the PAL System IFU “IM-PAL-650”.