Chapter 5: Troubleshooting
5-LV-1-4 8/3/11 LV1 User Manual 5-4
Process pressure builds
but does not extend
Plugged interaction chamber (IXC)
CAUTION: RELIEVE
PRESSURE BY SWITCHING THE MODE
SELECTOR TO THE OFF POSITION
PRIOR TO TRYING ANY REMEDY.
REVERSE THE POSITION OF THE IXC.
RUN THE UNIT UNTIL THE PRODUCT
FLOWS. RETURN PARTS TO THEIR
ORIGINAL POSITION. RUN
APPROPRIATE SOLVENT THROUGH
SYSTEM.
Short process piston
seal life
Check compatibility charts
Ensure proper installation
Damaged bearings, piston or bore of liquid end
body
Replace bearings, piston or liquid end
body
Processes faster and
louder than normal and
does not develop full
pressure
Fill feed reservoir and re-prime the
system
Leaks at coned and
threaded connections
Refer to the Coned and Threaded
Connections Maintenance section of
Chapter 4.
Reverse flow through
check valve
Check inlet check valve for correct
assembly, leakage, and/or blockage
Ensure that the priming valve is shut
tight
Processes faster than
normal; equipment will
not develop full
pressure