7BP B3 AFIB
EN
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Accuracy test
We
recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
10. Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
• Batteries and parts that become worn with use are not included.
• Opening or altering the device invalidates the guarantee.
• The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
•
The cuff has a functional guarantee (bladder tightness) for
2
years.
Please contact your local Microlife-Service (see foreword).
11. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
402
g (including batteries)
Dimensions:
138 x 94.5 x 62.5
mm
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280
mmHg – blood pressure
Cuff pressure
display range:
Resolution:
Static accuracy:
Pulse accuracy:
Voltage source:
Battery lifetime:
IP20:
Reference to
standards:
40 - 200 beats per minute – pulse
0 - 299 mmHg
1 mmHg
pressure within ± 3 mmHg
± 5 %
of the readout value
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional
approx. 920 measurements
(using new batteries)
Protected against solid foreign particles
with a diameter of more than 12.5 mm,
no protection against water
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Clean the cuff by using a clean cloth with clear water or 1%
condensed bleach, then stay at shady & cool place until natural
air dry.
WARNING: Do not wash the cuff in a washing machine
or dishwasher!