Rotary Microtome HM 340 E
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf 386 200 - English
2-8 TURNING OFF THE FUNCTION
<RETRACTION>
If desired, the function <retraction> can be turned
off. This can be achieved by pressing the buttons
(fig. 6.5 and 6.8) on condition that the lower line
of the display (fig. 6.1) is blank.
The letter R (last line on the right side on the
display) does not light up anymore.
The yellow LED RETRACTION, however, might
still be on and goes off only after having passed
from the return travel again to the cutting
movement by means of the handwheel.
To turn on again the function <retraction>, press
button (fig. 6.8) together with button (fig. 6.5),
while the lower line in the display is blank.
2-9 SELECTING THE INDICATED
LANGUAGE
The sectioning information on the display can be
shown alternatively in three languages. English,
German and French are available.
The instrument must be turned off. To select one
of these languages, press buttons (fig. 6.7 and 6.8)
simultaneously. Then turn on the power switch of
the instrument with still pressed buttons (fig. 6.7
and 6.8). Then LANGUAGE DEUTSCH lights up
on the display (fig. 6.1). To select another
language, press button (fig. 6.7) once or twice until
the desired language is shown. To confirm the
desired language, press the button (fig. 6.8) and
turn off the instrument on the power switch. Wait
a few seconds and then turn on the instrument
again. The display (fig. 6.1) now shows the terms
in the desired language.
If the originally desired language was skipped
over, turn off the instrument and repeat above
described process.
If the button (fig. 6.7) is pressed more than twice
passing the language selection process, the user
has entered the service-routine-program. This
program can only be operated by trained service
technicians. To terminate this program, turn off the
power switch of the instrument.