PART 5 MAINTENANCE
Annual routine maintenance
To secure section quality and functioning of the
microtome, it is recommended that a routine
maintenance be performed by a trained service
technician once a year.
Service contract
MICROM offers a service contract which
guarantees that your instrument is always in
perfect condition. For more information, please
contact the nearest MICROM sales office.
Exchange of fuses
The two mains fuses are installed on the back of
the instrument beside the power switch (fig. 26.1).
To exchange the two fuses, turn off the power
switch of the instrument and unplug the
instrument. Open the small cover (fig. 26.2) of the
fuse holders from the right side by means of a flat
screw driver. Pull out the fuse holders, each of
which is marked with an arrow, and exchange the
fuses against new ones.
Rating of fuses:
For power requirements: 220-240 V:
- 2 fuses 0,4 AT slow-blow
For power requirements: 100-120 V:
- 2 fuses 0,8 AT slow-blow
Fig. 26
Caution:
Any shipping of the instrument requires original
packaging materials! Damages caused by shipping
with non-conform packaging are not covered by the
manufacturer warranty! Any damage repairs
resulting in non-conforming package are fully
charged to the sending party. We reserve the right
depending on seriousness of damage NOT to
repair. To order original packaging materials,
please contact MICROM International or your local,
by MICROM International authorized, dealer.
Caution:
The user must care for a clean and safe
condition of the instrument when returning it to
an appropriate service provider.
Note:
If the original packaging is no longer available,
please contact your local MICROM representation.