Rotary Microtome HM 340 E
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf 386 200 - English
Installing the magnifier on the microtome
Please use the enclosed fastening elements to
attach the magnifier to the microtome. Remove the
front plastic caps from the upper side of the
housing.
If there are no drilled holes in the housing, it must
be replaced with a new one with drilled holes.
Fasten the peg (fig. 25.4) to the plate (fig. 25.6)
with the screw (fig. 25.7). Carefully introduce the
four sleeves (fig. 25.8) into the drilled holes with
the screws (fig. 25.5).
CAUTION! Please note that the sleeves do not
fall into the interior of the microtome.
CAUTION! Make sure the plate (fig. 25.6) is
mounted in a way that the peg (fig. 25.4) is closer
to the front of the microtome.
Put the plate (fig. 25.6) on the sleeves and fasten it
to the microtome with four screws (fig. 25.5).
A movable plastic socket that should be put on the
peg (fig. 25.4) is on the under side of the
magnifier. The magnifier can be adjusted forwards
and backwards for the most favourable viewing
position.
If the large field magnifier is not used, slightly
raise it and turn it sideways on the peg.