48
tions properly.
After checking all of the conditions above, repeat the
quality control test with a new test strip. If the quality
control test results are still outside of the range printed
on the test vial (or foil pouch), there may be a problem
with your meter. Please seek help and contact your
dealer.
Control Solution 1 is sufficient for most self testing
needs. If you think your meter or strips may not be
working correctly, you may also want to do a level 2
test. The ranges for both (CTRL 1 and CTRL 2) are
displayed on the test strip vial (or on the foil pouch).
Simply repeat the test procedure, using Control Solu-
tion 2. For conrmation of results, Control Solution 1
tests should fall within the CTRL 1 range, and Control
Solution 2 tests should fall within the CTRL 2 range.