47
of the screen. If the result falls within the range print-
ed on the test strip package (typically CTRL 1) the
device is working properly.
After the test is finished, eject the test strip. If the
control solution results are outside of the reference
range:
- Conrm you are matching the correct range. Con-
trol Solution 1 results should be matched to the
CTRL1 range printed on the test strip vial (or foil
pouch).
- Check the expiration date of the test strip and
control solution. Make sure that the packages have
not been opened for more than 6 months. Discard
any expired test strips and control solutions.
- Confirm that you are testing within the correct
temperature range (15-30 °C). Make sure that the
test strip vial and control solution bottle have been
tightly closed.
- Make sure that you are using the correct brand of
control solution.
Make sure you are following the user guide instruc-