MILESTONE
USER MANUAL
Ethos EZ User manual MA073-006 Sensors and Accessories chapter -33-
WORKING PRESSURE LIMIT
Rotor P limit*
HPR (SK-10) 45
MPR (SK-12) 30
PRO-16/24 25
* at 200°C
QUALITY PRESSURE SENSOR (QP) (option)
QP quality pressure sensor will detect the presence of reaction products (vapors), QP= acids or
QPS= solvents, in the cavity and a system response when the concentration of such products rises
above a defined limit.
This sensor prevents the venting of the vessel reducing the power in accordance with the set limit.
HOW TO SET THE QP SENSOR
Enter in the software and select the “Parameter” page:
Clicking in the QP percentage window select the QP limit: