LOGOS One
All-in-one Rapid Tissue Processor
Operator Manual
MM098
MILESTONE
HELPING
PATIENTS

Table of Contents

Questions and Answers

  • B
    Brittany SchneiderAug 12, 2025
    Why do I see fatty tissue collapse in blocks & microsections with Milestone LOGOS One?
    • R
      rcookAug 12, 2025
      If you observe fatty tissue collapse in blocks and microsections, it might be due to inadequate tissue fixation reacting with JFC, or the absence of connective tissue parenchyma. To address this, fix the tissue conventionally overnight or include heated fixation. Alternatively, use the JFC followed by Isopropanol protocol for processing.
  • J
    jerry74Aug 16, 2025
    What to do if there is incomplete clearing of fat in Milestone LOGOS One?
    • M
      Miss Chelsea WhiteAug 16, 2025
      Incomplete clearing of fat can occur if the processing cycle is too short or if isopropanol is inefficient for fat clearing, or if the dehydration/clearing agent reuse life has been exceeded. Using an appropriate size processing cycle or reducing the tissue block thickness may help. As an alternative clearing agent, use JFC, or replace the dehydration or clearing reagent.
  • J
    Justin WangAug 19, 2025
    Why are there pigment deposits in sections when using Milestone LOGOS One?
    • H
      Heather KellerAug 19, 2025
      Pigment deposits in sections may arise from incomplete fixation of blood. Perform a pigment test to determine if they are formalin pigment or hemosiderin deposits. Consider including heated fixation or extending conventional fixation time to improve fixation.
  • M
    Matthew SnyderAug 23, 2025
    What to do if specimens detach from surrounding WAX in Milestone LOGOS One Laboratory Equipment?
    • G
      Grant OdomAug 23, 2025
      Specimens detaching from the surrounding wax may be due to a processing cycle that is too short. Try using an appropriate size processing cycle or reducing the tissue block thickness.
  • S
    Stacy CruzAug 27, 2025
    How to fix white appearance of block face with Milestone LOGOS One Laboratory Equipment?
    • P
      Paula KingAug 27, 2025
      A white appearance of the block face indicates that wax infiltration is not entirely complete. To resolve this, increase the wax infiltration time.
  • K
    Kyle OrrAug 31, 2025
    What causes collapse of block face after sectioning with Milestone Laboratory Equipment?
    • K
      kathleenchavezAug 31, 2025
      Collapse of the block face after sectioning suggests that wax infiltration is not entirely complete. Increasing the wax infiltration time should resolve this issue.
  • T
    Troy JohnsonSep 4, 2025
    How to resolve poor dehydration in Milestone Laboratory Equipment?
    • C
      connie77Sep 4, 2025
      Poor dehydration can be caused by alcohol contamination (water), the use of sponge pads or inefficient biopsy pads, or a processing cycle that is too short. Replace the alcohol, extend the processing time, use sponge alternatives or use appropriate size processing cycle to resolve this.
  • G
    garnerkellySep 7, 2025
    What causes pyknotic nuclei and severe cytoplasmic distortion in Milestone Laboratory Equipment?
    • C
      Charles DurhamSep 8, 2025
      Pyknotic nuclei and severe cytoplasmic distortion can result from poor fixation. Extend the fixation time, or use the JFC/Isopropanol program to address poor fixation or JFC reaction issues.

Summary

Introduction and Specifications

Symbol Usage and Warnings

Explains warning and informational symbols used throughout the manual.

Installation and Initial Setup

Space, Electrical, and Power Requirements

Physical space, power, and electrical connection guidelines.

Operating the LOGOS ONE

Running Processing Programs

How to start and manage various processing modes.

Reagents and Wax Management

Reagent Reuse Life and Settings

Managing reagent expiration, cycles, and usage limits.

Reagent and Wax Replacement Procedures

Step-by-step guide for replacing reagents and wax.

Safety, Alarms, and Error Handling

Power Failure and Loading Time-Out Procedures

How the system behaves during power interruptions.

Maintenance and Support

Troubleshooting Guide

Tissue Processing Issues Overview

Common problems and solutions for tissue processing.

Milestone LOGOS One Specifications