MM098-002 – LOGOS ONE – Operator Manual
113
9.3 Beware
The use of inserts like thick paper or the use of small mesh cassette negatively influences the correct flow
of the reagent through the cassette. Use only histology papers or cassette with large mesh to allow proper
unobstructed flow of reagent.
In the case sponges are used, increasing of the Ethanol, JFC Isopropanol and WAX
processing time by half of the preset time is recommended.
Use ONLY Histology sponges.
If the Laboratory is already using particular kind of cassettes, please contact either our
application support: application@milestonemedsrl.com with details of the type of
cassettes utilized, so that we can check whether they are suitable to be used in our units.
Histology paper – Biopsy large mesh
cassette
YES
Thick paper – small mesh cassette
NO