EasyManua.ls Logo

Milestone LOGOS One - Maintenance and Support; Routine and Periodic Maintenance Tasks

Milestone LOGOS One
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Milestone s.r.l.
MM098-002 LOGOS ONE Operator Manual
102
7 MAINTENANCE
7.1 Introduction
Maintenance of LOGOS ONE is carried out at two levels:
The first is at laboratory level: a simple routine maintenance can prevent the likelihood of more serious
problems. Minimal routine maintenance is required by laboratory personnel that use LOGOS ONE. The
following procedure is recommended to ensure optimal performance of LOGOS ONE:
The second level of maintenance is technically more detailed and is beyond the scope of normal laboratory
personnel. The unit requires once a year a technical preventive maintenance to maintain the high
performance and prevent any malfunctionality.
Contact annually your local distributor or Milestone offices directly
(customersupport@milestonemedsrl.com) how to perform Preventive Maintenance and
regular inspections in order to maintain your LOGOS ONE within safe operative standards.
7.2 After each run
Ensure that LOGOS ONE is generally kept clean after each use. Do not use abrasives or sharp instruments
that may damage external surface etc.
Clean as soon as possible the rack (by molten wax) after every processing cycle to completely remove the
wax. For more details see the paragraph 7.2.1.
Ensure that the wax level is between the MIN and the MAX (wax settings screen). For more details see the
paragraph 2.13.
Wipe off the undersurface of the cavities covers when the cassettes rack is unloaded from the wax, on
completion of both Processing and WAX/MAIN cavity cleaning procedures.
Traces of solidified wax may interfere with correct closing of the cavity and may result in
vacuum error.
7.2.1 Rack cleaning procedure
There are two different ways to correctly clean wax traces away from the LOGOS ONE RACK after
processing:
FIRST WAY:
It is possible to clean the rack inside the processing chamber during a Cleaning cycle with 3 tanks (see the
chapter 3.3), after a wax phase of an automatic processing, when the wax has been loaded in the main
chamber from the wax chamber.

Table of Contents

Related product manuals