EasyManua.ls Logo

Milestone LOGOS One - Tissue Processing Selection Guidelines

Milestone LOGOS One
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Milestone s.r.l.
MM098-002 LOGOS ONE Operator Manual
115
9.5 Tissue processing selection guidelines
Ethanol-Isopropanol-WAX
1. Ethanol-Isopropanol recommended as the standard routine method for small to medium-large size
biopsies.
2. Fat content of the tissue is kept to a minimum (< 2mm thickness).
3. Very Cost effective for general tissue-processing.
4. Strongly recommended for processing of Renal/Gastric, breast core biopsies and equivalent tissue
size.
Eth-JFC-Isopropanol-WAX/ JFC-Isopropanol
1. JFC-ISO sequence is strongly recommended for high fat content such as Breast / Lipoma up to full
size cassette in less than 4 hours.
PROWAVE-WAX
2. PROWAVE is a “one step” solution that can be used for tissues where overnight fixation time is
uncertain, since PROWAVE can also act as a secondary fixative.
3. Ideally suited for non-fatty tissues (up to 5mm thick) or moderate lipid content tissues (up to 3mm
thick).
Ethanol-Xylene-Isopropanol-WAX
1. Ethanol-Xylene is not recommended from Milestone, is up to the user to decide the use of xylene. It
can be used as the standard routine method for small to medium-large size biopsies.
Ethanol-Isoparaffin (MileGREEN®)-Isopropanol-WAX
1. Isoparaffin (MileGREEN®)-ISO sequence is recommended for high fat content such as Breast /
Lipoma.
Refer to chapter 3.3.1 Tanks for process and main cavity cleaning, for suggestions on how
to set up tanks in LOGOS ONE.