MM098-002 LOGOS ONE Operator Manual
9
1.5 Warning information
Should the equipment be used in different ways than those specified by the manufacturer, the warranty
provided by the manufacturer may be void.
Main power plugs are considered disconnecting device. Disconnect all plugs from the socket outlet before
assembling of the system, before connection of the accessories and before cleaning.
Equipment shall be connected to a main power socket outlet which is accessible and visible by the operator.
Equipment is provided with replaceable fuses, however their replacement is to be performed by properly
authorized service personnel. If the system is not properly working, please contact the local authorized and
trained service representative of the manufacturer.
All parts of the equipment and all accessories must be supplied only by the manufacturer.
DO NOT OPEN THE CHASSIS. Failure to follow the instruction in the documentation
might lead to a reduction of device safety. If the equipment is not working properly,
please contact your supplier or the manufacturer.
Do not remove the instrument, accessories, panels or cover. Only authorized and
qualified SERVICE personnel may repair the instrument and access the instrument’s
internal components.
It is important that normal standards for safety and good laboratory practices are
applied. Always use common sense and the best known practice when operating the
instrument. Failure to follow the instructions in the documentation might lead to a
reduction of device safety.
The Competent Authority owning the instrument has primary responsibility for
accident-free operation together with designated personnel who operates SERVICE
or repair it.
To avoid damage to the instrument, use only the reagents listed in section 1.7
Compatible reagents. Failure to follow the instructions in the documentation might
lead to a reduction of device safety.
Use caution when handling hot, molten WAX.
METAL PARTS
Never introduce metal parts into the MAIN CAVITY.
High voltage discharges may cause uncontrolled increases of temperature, which
could damage the unit and result in incorrect tissue processing also. Failure to
follow the instructions in the documentation might lead to a reduction of device
safety.
Be careful: exhaust reagents are toxic and carcinogenic.
Handle with care and follow the local standard regulations.