Requirements and specications
40
Systemspecications
Electrical
The power supply converts mains voltage to 28 V. The power supply is compatible worldwide.
Catalog Frequency Max power Voltage
ZIQ7003T0 50-60 Hz ±10% 600 VA 100 – 240 V ± 10%
ZIQ7005T0 50-60 Hz ±10% 600 VA 100 – 240 V ± 10%
ZIQ7010T0 50-60 Hz ±10% 600 VA 100 – 240 V ± 10%
ZIQ7015T0 50-60 Hz ±10% 600 VA 100 – 240 V ± 10%
Environmental
Specic environmental conditions have to be respected to ensure normal operation of the Milli-Q
system.
Location indoor use only
Ambient operating temperature room temperature ranges from 4°C to 40°C
Relative humidity
room relative humidity conditions of 80% for a
temperature up to 31°C, decreasing linearly to a
relative humidity of 50% at 40°C
Altitude up to 3000m above sea level
Installation category bench / underbench or wall mounted
Pollution degree 2
Noise level < 50 dB at 1m