Requirements and specications
36
REQUIREMENTS AND SPECIFICATIONS
Waterspecications
Elix Water
The system has been designed to produce water according to specications when operating
within feed water requirements.
Parameter Value or range
Resistivity > 5 MΩ.cm @ 25°C
Conductivity < 0.2 μS/cm @ 25°C
Total Organic Carbon (TOC)
≤ 30 ppb
Particles (size > 0.22µm)
No particles with size > 0,22 µm if 0,22µm lter used
Bacteria
< 0,01 CFU/mL (with Millipak® 40 or Biopak® lters when
installed and used in a laminar ow hood)
Pyrogens (endotoxins)
< 0,001 EU/mL (with Biopak® lter when installed and used in
a laminar ow hood)
RNases < 1 pg/mL (with Biopak
®
lter)
DNases < 5 pg/mL (with Biopak
®
lter)
Proteases < 0.15 µg/mL
Flow rate
3 L/h for IQ 7003
5 L/h for IQ 7005
10 L/h for IQ 7010
15 L/h for IQ 7015
The Milli-Q IQ 7003/7005/7010/7015 system is intended to produce ultrapure water that
meets or exceeds requirements as described by the organizations below:
Organization Water quality / grade
European Pharmacopeia Puried water
U.S. Pharmacopeia Puried Water
Japanese Pharmacopeia Puried Water
Chinese Pharmacopeia Puried Water
ASTM
®
D1193 Type II water Grade B
ISO
®
3696 Grade 2 water
Chinese National Standard GB 6682 Grade 2 water
JIS K 0557 A2 water