EasyManua.ls Logo

Millipore Pellicon Single Pass TFF - Page 8

Millipore Pellicon Single Pass TFF
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 www.sigma-aldrich.com
1. Loosen the nuts on the tie rods of the holder and remove the
nuts, washers, tie rod spacers, and end plate.
2. Slide the Pellicon
®
2 Mini Cassette gasket onto the holder so
that the tie rods pass through the cutouts in the gasket.
3. Slide the Pellicon
®
Single-Pass TFF mini diverter plate onto
the holder so that the tie rods pass through the cutouts in the
diverter plate. Note the location of the fittings on the diverter
plate.
4. Slide a gasket onto the holder.
5. Slide a Pellicon
®
2 Mini Cassette onto the holder.
6. Slide a gasket onto the holder.
7. Slide a diverter plate onto the holder so that the fittings on
the diverter plate are in the opposite location of the previous
diverter plate.
8. Slide a gasket onto the holder.
9. Install the next Pellicon
®
2 Mini Cassette.
10. Slide a gasket onto the holder.
11. Complete the installation with the required number of diverter
plates, alternating the location of the ports on the diverter
plates.
12. Slide a gasket onto the holder.
13. Install the end plate of the holder and torque the holder
using the specifications described in the Pellicon
®
2 Cassette
Installation Guide.
14. To nalize the serial ow path, cap or connect tubing to
the diverter plates ports (female luer lock) according to the
descriptions listed in the table in this section.
NOTE
For details on installation and torque specification, please
refer to the Pellicon
®
2 Cassette Installation Guide.