EasyManua.ls Logo

Millipore SDS 200 - Electrical Specifications; Environmental Specifications; Hydraulic Specifications

Millipore SDS 200
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Information
SDS 200 - SDS 350 7
ELECTRICAL SPECIFICATIONS
Voltage / Frequency / Power Cat. Number Main Fuse
100V ±10%/50-60 Hz/800 VA with pump
TANK7P200
TANK7P350
10 A T fuse
Millipore SAS Spare Part FTPF04803
100V ±10%/50-60 Hz/10 VA without pump
TANK70200
TANK70350
1 A F fuse
Millipore SAS Spare Part FTPF00756
120V ±10%/60Hz/800 VA with pump
TANK6P200
TANK6P350
10 A T fuse
Millipore SAS Spare Part FTPF04803
120V ±10%/60Hz/10 VA without pump
TANK60200
TANK60350
1 A F fuse
Millipore SAS Spare Part FTPF00756
230V ±10%/50Hz/800 VA with pump
TANK5P200
TANK5P350
3.15 A T fuse
Millipore SAS Spare Part FTPF03213
230V ±10%/50Hz/10 VA without pump
TANK50200
TANK50350
0.5 A T fuse
Millipore SAS Spare Part FTPF00306
The source of electrical power must be earth grounded.
ENVIRONMENTAL SPECIFICATIONS
Indoor use only.
Ambient storage temperature: 5°C< T <40°C
Altitude: < 3000 m
Installation Category: II
Pollution Degree: 2
HYDRAULIC SPECIFICATIONS
Purified water inlet to reservoir: 8 mm OD tubing
Distribution loop: inlet and return ¾ inch NPT Male
Drain capacity (overflow): 220 LPH at a maximum height of 200 mm from floor level
Maximum length of drain tubing: 3 m