APPENDIX 1 MAIN FUSE REPLACEMENT
1. Switch off power to the system. Switch main power ON/OFF switch OFF.
2. Unplug the power cord from the power supply.
3. Unscrew the cap on the fuse holder "FUSE" on the front panel.
4. Pull out the faulty fuse and replace it with a new one.
Voltage Cat. Number Main Fuse
100V ±10% with pump
TANK7P200
TANK7P350
10 A T fuse
Millipore SAS Spare Part FTPF04803
100V ±10% without pump
TANK70200
TANK70350
1 A F fuse
Millipore SAS Spare Part FTPF00756
115V ±10% with pump
TANK6P200
TANK6P350
10 A T fuse
Millipore SAS Spare Part FTPF04803
115V ±10% without pump
TANK60200
TANK60350
1 A F fuse
Millipore SAS Spare Part FTPF00756
230V ±10% with pump
TANK5P200
TANK5P350
3.15 A T fuse
Millipore SAS Spare Part FTPF03213
230V ±10% without pump
TANK50200
TANK50350
0.5 A T fuse
Millipore SAS Spare Part FTPF00306