EasyManua.ls Logo

Milwaukee MC510 - Operating the ORP Meter; Step-by-Step Operation Guide; Setpoint Configuration; Adjusting the Setpoint Value

Milwaukee MC510
2 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION:
Connect the supplied 12
VDC power adapter to the
meter and to the mains line.
Connect the ORP electrode
to the BNC socket on the
bottom of the meter.
Always remove the elec-
trode protective cap before
taking any measurement. If
the electrode has been left
dry, soak the tip (bottom 2.5
cm) in rinse solution
(M10000B) for a few minutes to reactivate it.
Turn the instrument on by pressing the ON/OFF key.
Allow the reading to stabilize and the meter will start continuous
monitoring.
The meter is factory calibrated.
A blinking alarm will indicate when the measured ORP value is lower or
higher than the setpoint, depending on user selection.
NOTE:
The power contact is not protected by fuse inside the meter. It is
recommended to protect it outside, against failure.
WARNING: when the alarm led is blinking, the mains voltage is
present in the output power socket.
SETPOINT:
The setpoint can be selected by
adjusting the central front knob to the
desired value.
The selectable range for MC510 is
from 0 to 600 mV.
The nature of the setpoint can be selected by setting the switch to the
desired position (ABOVE or BELOW).
INSTALLATION PROCEDURE:
Test Equipment Depot - 800.517.8431
- 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Related product manuals