Part II General Information
Structure and Assembly/Disassembly 181
3. Remove the screws fixing the bottom case to the keyboard cover.
Use a crosshead screwdriver to remove the 10 M3X8 Nylok screws on the bottom case.
4. Remove the FPC flat cable between the main board and the control panel.
Lift the keyboard cover softly. When the FPC flat cable between the main board and the control
panel is visible, remove the flat cable from the interface on the main board to detach the keyboard
cover.
LCD signal cable
M3X8 Nylok
screw