Probes and Biopsy 13-13
Low-level disinfectant
EPA
Registration No.
Name
9480-6
SANI-CLOTH PLUS GERMICIDAL DISPOSABLE CLOTH
70627-56
OXIVIR TB
9480-9
PDI Sani-Cloth AF3 Germicidal Disposable Wipe
46781-8
CAVIWIPES
46781-6
CAVICIDE
13.1.6 Environment
Working condition
Use the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
L12-3E 0~40
30%85%RH (no
condensation)
700hPa~1060hPa
C11-3E 0~40
30%85%RH (no
condensation)
700hPa~1060hPa
SP5-1E 0~40
20%85%RH (no
condensation)
700hPa~1060hPa
DE11-3E 18~30
20%85%RH (no
condensation)
700hPa~1060hPa
SC6-1E 0~40
20%85%RH (no
condensation)
700hPa~1060hPa
V11-3HE 0~40
20%85%RH (no
condensation)
700hPa~1060hPa
L9-3E
0~40
20%85%RH (no
condensation)
700hPa~1060hPa
SD8-1E
18~30
20%85%RH (no
condensation)
700hPa~1060hPa
L14-5WE
0~40
20%~85%RH (no
condensation)
700 hPa ~ 1060 hPa
P7-3E 10~40
30%85%RH (no
condensation)
700hPa~1060hPa
L20-5E 0~35
15%~80%RH (no
condensation)
700 hPa ~ 1060 hPa
SC5-1NE
0~40
20%85%RH (no
condensation)
700hPa~1060hPa

Table of Contents