Mindray DC-90 - Page 313

Mindray DC-90
391 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Probes and Biopsy 13-33
(4) Hold the probe and press the pressing block to separate V-shaped guiding block from the
pressing block.
(5) Insert the needle into V-shaped guided block.
(6) Release the pressing block, adjust the angle adjusting nut to confirm that the needle can
freely slide in a vertical direction.