Part I Service Scenarios
Repair 97
3.2.3.6 Troubleshooting Flow
White stripe & black stripe
Replace the probe
No
Does t he problem
rec ur?
Replace the probe
No
Replace the main board
Yes
Does t he faul t r ecur?
Yes
Replace the probe
connection board
End
No
Replace the probe or probe
connection board with
obvious di fferences
Yes
Is the fault rectified?
Yes
No
Is the fault rectified?
No
Yes
Is the fault rectified?
Yes
Seek support from R&D
personnel
No
Are the p ins o f the
pr obe plug (two ends)
abn ormal ob vious ly?